← Components & BoP
Downstream

Instrumentation and safety

Gas quality, dew point and flow measurement across the whole plant

Running an electrolyser means measuring gas quality, pressure, flow and leaks. We start by setting out what has to be measured where, then confirm range, accuracy, signal type and the certification required in your country before finding a model.

Control basis
2%below · H₂ in O₂ · the limit set by Korean and international safety regulation
Understand

What this part does and what decides it

Why it matters

Run an electrolyser and a little hydrogen crosses over into the oxygen. The hydrogen mixed into that oxygen must stay below 2 percent, which is why it has to be watched continuously. That is not a figure we set: Korean and international safety regulation both work to it. There is one mistake people make when choosing the instrument: setting the measuring range from the value at full output. This ratio rises the lower the current density you run at and the higher the pressure. Measured on our own stack, the lowest and highest points are 4.5 times apart. So there are only two things we need from you: how far down you will turn the output, and how high you will take the pressure. We size the range and set the alarm points against those two. Look at the running cost early as well. Some sensors are consumed in use and some are not. If you have to read very low concentrations, you end up on the consumable kind, and the replacement cells keep costing. Put that in the budget from the start.

Background

Why the ratio rises as you turn the load down

The amount of hydrogen crossing the membrane is nearly independent of current. It crosses because of a pressure difference, so much the same amount comes through even when you pass less current. Oxygen, by contrast, is produced in proportion to current: drop the current to 40 percent and the oxygen drops to 40 percent too. On the same stack, lowering the current is the same as lowering the current density — that is the figure quoted in A/cm² in the data. What is managed is the ratio of hydrogen to oxygen. The hydrogen coming across stays as it was while the oxygen falls, so the ratio rises. The same hydrogen is mixing into less oxygen. Pressure raises it by a different route. Higher pressure widens the hydrogen pressure difference across the membrane, so more hydrogen crosses in the first place. Low load and high pressure together bring both effects at once. The measured figures: on a 23-cell HXS-2 (0.1 M KOH, 43–45 °C), 0.39 percent at rated load and ambient pressure, 0.66 percent at 40 percent load and ambient, 0.98 percent at rated load and 6 barg, and 1.75 percent at 40 percent load and 6 barg. Lowest to highest is a factor of 4.5, and at 6 barg with 40 percent load only 0.25 percentage points remain to the 2 percent control limit.

What each measuring principle actually measures

The Measuring principle column says what the instrument physically responds to. For the same oxygen concentration, each principle is strong over a different range and carries different consumables and calibration intervals. Read this column first and it becomes clear why some models belong at trace level and others at percent level. For concentration there are two families. Principles that consume the target gas reach lower, but the sensor has a service life. Principles that do not consume it last longer and hold calibration longer, but lose resolution as the concentration falls. Flow meters and pressure instruments sit outside those two families: they measure how much is flowing, or a pressure, rather than what the gas is made of. On pressure the table covers the ones that put out an electrical signal (piezoresistive, ceramic capacitive), the mechanical ones you read off a pointer without power (Bourdon tube, diaphragm), and electromechanical contacts that make or break the circuit themselves. How far down a given model reaches varies within a principle. That is what the Range column in the table records. Infrared (NDIR), the method most people reach for first in leak detection, cannot be used on hydrogen. The hydrogen molecule does not absorb infrared, so there is no signal to read. Do not move an infrared detector used for another flammable gas onto a hydrogen duty. Hydrogen leaks are caught with catalytic bead or electrochemical sensors. Leak detection splits two ways. A fixed detector is mounted on a wall or a post to watch the concentration in the area continuously, and it puts out 4-20 mA or a contact so it can drive an alarm and an interlock. A portable one is carried to find where a leak you already know about is coming from. Neither substitutes for the other, which is why the table is split on that axis: a portable unit left in place sees nothing while nobody is holding it, and a fixed one cannot be walked along a run of fittings joint by joint.

Electrochemical, fuel-cell type
Measures the current produced as the target gas reacts at an electrode; the current is proportional to concentration. The cell generates its own current like a fuel cell, so no external supply is needed. Strong at low levels, reaching ppm and, on some models, below. The electrode and electrolyte are consumed by the reaction, so the cell is a replacement item
Polymer electrochemical
A solid polymer membrane in place of a liquid electrolyte. Nothing to dry out or leak, so there is no top-up and no restriction on mounting orientation
Thermal conductivity
Uses the fact that gases carry heat at different rates: the heat leaving a warmed element reads the composition. Hydrogen conducts about seven times as well as air, so the contrast is large and this principle suits it particularly well. No consumables, but it only works on a two-component mixture; a third gas leaves the reading ambiguous
Thermo-paramagnetic
Oxygen is drawn into a magnetic field and other gases are barely affected. The strength of that pull gives the oxygen concentration. It does not consume oxygen, so it is long-lived and strong at percent-level concentrations
Optical luminescence
A fluorescent layer emits light when illuminated, and oxygen reaching it suppresses that emission. Concentration is read from how far the emission drops and how long it takes to recover. Oxygen is not consumed, so calibration holds for longer; the fluorescent cap is the periodic replacement item
Thermal mass flow
Measures the heat a flowing gas carries away from a heated element. The correction factor differs by gas, so the unit has to arrive configured for hydrogen
Laminar differential pressure mass flow
Uses the fact that the pressure drop across a laminar element is proportional to flow. Pressure and temperature are measured alongside it and converted to mass flow. Response is fast
Catalytic bead (pellistor)
A flammable gas burns on a catalyst-coated bead and the heat raises the resistance of a platinum coil, which is what is measured. It is read as a percentage of the lower explosive limit and used to watch a room for leaks. Combustion has to happen, so a minimum oxygen concentration is required; in oxygen-poor or inert atmospheres the reading runs low. Catalyst poisoning sets the life
Tracer-gas sniffing
The part under test is filled with a hydrogen tracer gas, usually 5% hydrogen in 95% nitrogen, and a probe carrying a sensor that responds only to hydrogen is passed over the surface to find where it escapes. The scale is ppm rather than %LEL, so it catches very small holes and the probe pinpoints where they are. The gas has to be put into the part, so it cannot watch the concentration in a room around running equipment
Piezoresistive
Pressure deflects a diaphragm and changes the resistance of a silicon chip behind it. Wide span and low cost make it the default for pressure. The wetted face is metal, so which metal it is decides media compatibility
Capacitive (ceramic)
The capacitance between a ceramic diaphragm and a backplate changes as the diaphragm deflects. The wetted face is ceramic rather than metal and there is no fill fluid. This is why it is the one to look at instead of a metal diaphragm for alkaline and acidic media
Bourdon tube (mechanical)
Pressure straightens a curved tube and that movement drives a pointer. It reads without power, which is what you want for commissioning and walk-round checks, but it sends no signal, so it cannot serve logging or an interlock
Diaphragm (mechanical)
A thin diaphragm deflects under pressure and drives the pointer. It holds lower pressures better than a Bourdon tube, and a PTFE lining or a special-material pressure chamber makes it fit corrosive media
Electromechanical contacts
Above a set pressure a spring and lever physically make or break a contact. Nothing electronic sits in the path, so it still acts when the controller is down — which is why it is used to wire a shutdown interlock independently

How to read a hazardous-area marking

If the instrument goes into a hazardous area it has to be rated for that area. A single line such as Ex db IIB+H₂ T6 Gb in a catalogue is four separate things written end to end. The db at the front is the protection method. Flameproof (the Ex d family) contains an ignition inside a strong enclosure so it cannot spread; intrinsic safety (Ex ia, Ex ib) keeps the energy in the circuit below what it takes to ignite anything in the first place. Analysers with a display and mains power are usually flameproof; small transmitters with no display are usually intrinsically safe. The IIB or IIC in the middle is the gas group. The further along it goes the easier the gas is to ignite, and hydrogen sits in the most demanding group, IIC. A unit certified only to IIB cannot be used in a hydrogen area. Some markings add hydrogen on its own as IIB+H₂, so do not rule a unit out on the letters IIB alone. T6 is the temperature class. It says the surface stays below the ignition temperature of the gas, and T1 through T6 guarantees progressively lower surface temperatures. The Gb at the end is the equipment protection level: Ga corresponds to Zone 0, Gb to Zone 1, Gc to Zone 2. The certification schemes differ by country: ATEX in Europe, IECEx internationally, KCs in Korea. The same unit may still need a separate KCs certificate to be installed in Korea, so tell us the country of installation first. Below is only what we have confirmed against manufacturer documentation. A model left out is not a model without a hazardous-area version; it is one we have not confirmed. Note also that the hazardous-area version is usually a separate model number rather than an option on the same one. Get the part number wrong on the order and a general-purpose unit arrives.

Michell XTC601
Flameproof Ex db IIB+H₂ T6 Gb; ATEX, IECEx, KCs
Michell XTP601
Three build codes: general purpose GP1 and GP2, hazardous area EX1. EX1 is flameproof Ex db IIB+H₂ T6 Gb, dust Ex tb IIIC T85°C Db (IP66); ATEX, UKCA, IECEx, cQPSus, TR CU Ex
Messkonzept FTC 320
Non-flameproof; FTC 320 EX is a separate model
Chang AI CI-PC551-2
Flameproof Ex d IIB T6 Gb, IP65. The manufacturer manual prints the same field as Ex d C T6 Gb
GASDNA DA-600S
Flameproof Ex db IIC T6 Gb (IP6X), dust Ex tb IIIC T85°C Db; KCs, ATEX, IECEx, JPEx, CE
Southland OMD-507
Non-Ex as standard. An intrinsically safe option exists, but the manufacturer rates it for Class I Div 2, which is not the Zone 1 flameproof grade this column means
Michell Easidew
Easidew I.S. is the separate intrinsically safe model; IECEx, ATEX, UKCA
PRIGNITZ SPT
PMP-S122-Exi is the separate intrinsically safe part number; II 1G Ex ia IIC T4 Ga; CSA, ATEX, IECEx. Its datasheet states hydrogen approval
STORK DEWCom II
CE marking only in the manufacturer documentation; no hazardous-area rating confirmed
Bronkhorst EL-FLOW Select
EX-FLOW is the separate hazardous-area range
MSA ULTIMA X5000
ATEX, IECEx, CSA, FM, INMETRO and DNV-GL approvals; SIL 2 suitable. The marking depends on the part number and sensor
Honeywell Sensepoint XCD
ATEX II 2 GD Ex d IIC T6/T5 Gb, Ex tb IIIC T85°C Db; the same under IECEx. IIC gives hydrogen room to spare
Everything else
Models we have not yet confirmed against manufacturer documentation

Where you read the value

The same measurement is read in different places depending on the model. That is what the Display column records. A screen on the instrument is the simplest case. Where the sensor and the screen are one body it is integrated with the sensor; where it is built to drop into a cabinet door it is a panel-mount type. Either way somebody reads the number on site. A model without a screen only puts the value out as a current. That is the 4-20 mA signal, and the number is visible only if something at the other end receives it: a PLC, a recorder, or the manufacturer's own separate readout. Miss this and the instrument is fine while nobody can see what it reads. Not all 4-20 mA wiring is the same. A two-wire loop carries power and signal on the same pair; a three-wire unit takes its power separately. Connectors are standardised too, so a model using a DIN 43650 connector needs the matching socket ordered with it. Instruments with a wide span often add auto-ranging: the unit changes its own measuring range as the concentration moves. On a bare 4-20 mA link that means the same 20 mA is 1000 ppm at one moment and 100% at another. Decide whether you also want the signal that identifies the current range.

Sent is not the same as done

This is where instrumentation most often goes wrong against upper-level control. Whether the command reached the unit, whether the unit accepted it, whether it is running, whether it finished, whether it failed: each of these has to be a separate state. Without that separation you get a screen that reads normal while nothing is actually happening.

Agree units and scaling first

The same quantity comes in different units from different instruments. This table alone has oxygen-in-hydrogen reported in percent and in ppm. If the tag units are not matched and the 4-20 mA scaling is carried over as is, the reading is out by a factor of ten thousand, and it often sits in a plausible range long enough that nobody notices. If you also bring pressure in, settle barg against bara for the same reason.

Choose

The options and the models we handle

What you can choose

Start here and we adjust to your conditions. We can usually source what is not listed.

Safety monitoringBuilding your own stack or system setup

Continuous hydrogen concentration on the oxygen line. That is where the limit is reached first, so unattended operation starts here. The H2 in O2 group below is where these sit; catching a leak in the room itself is what the two Gas leak groups are for.

Quote this option
Hydrogen purityFuel cells, refuelling, analytical supply

Oxygen in hydrogen and dew point. These are the two values behind a stated purity target. Read the O2 in H2 and Dew point groups below together.

Quote this option
Hydrogen outputStack evaluation and data

Hydrogen and oxygen flow measured separately and compared with the theoretical rate from current. This is what gives you Faraday efficiency. The Gas flow group below is where these sit.

Quote this option
Pressure monitoringActually running the stack

Anode and cathode pressure read separately. Differential-pressure limits and interlocks take these as their input. The anode reads a KOH-laden fluid, so the wetted parts decide the model; the cathode reads saturated hydrogen gas, where hydrogen permeates metal, so the diaphragm is specified separately. The three pressure groups below — both sides of the stack, an independent interlock, and reading it on site — are where these sit.

Quote this option

Temperature is not on this page. It comes as one set with the heat exchanger, so we put it together on the electrolyte circulation side. Where each instrument goes is set out in the instrumentation table in the HXS-2 stack integration guide.

Models we can source

On oxygen in hydrogen, settle the order of magnitude first. Whether you manage it in ppm or percent is enough decides which principles are open to you — only electrochemical reaches ppm, and its cell is a consumable. The Range column is that order of magnitude.

Pressure is grouped by the kind of instrument. A transmitter sends a signal out, which is what logging, control and the differential-pressure calculation run on; a switch drops its contact mechanically at the set pressure, so it still acts when the controller is dead; a gauge is read by eye with no power and sends nothing. They are not alternatives to one another — a plant carries all three. For differential pressure we recommend calculating it from the two transmitters rather than adding a separate instrument.

Both gas tables carry a hazardous-area column. Those six are instruments we have bought and run ourselves, so every rating is confirmed against manufacturer documentation. The XTP601 is ordered either as a flameproof (EX) or a general purpose (GP) build, so its cell is marked optional. The other groups have no such column — only half of those models could be confirmed — and what we did confirm is listed under "How to read a hazardous-area marking" above.

H2 in O2

ModelMethodRangeDisplayHazardous areaQuote this
Michell InstrumentsXTC601EX1
Thermal conductivitySelectable from 0–1% up to 0–100%Built-in touchscreenFlameproofGet a quote
MesskonzeptFTC 320
Thermal conductivitySelectable from 0–0.8% up to 0–100%Built inNon-ExGet a quote
Thermal conductivityOrder specificBuilt in, capacitive touch keysFlameproofGet a quote
Thermal conductivityOrder specificBuilt-in 2.4-inch TFT LCDFlameproofGet a quote

O2 in H2

ModelMethodRangeDisplayHazardous areaQuote this
Michell InstrumentsXTP601
Thermo-paramagneticSelectable from 0–0.5% up to 0–50%Built inFlameproofoptionalGet a quote
Southland SensingOMD-507
Electrochemical, fuel-cell type0 to 10 ppm through 0 to 100%, selectableBacklit, auto-rangingNon-ExGet a quote

Dew point

ModelRangeDisplayQuote this
Michell InstrumentsEasidewEA2-TX
Dew point −100 to +20 °C, accuracy ±2 °CNone, 4-20 mA two-wireGet a quote
STORK InstrumentsDEWCom II
Dew point −100 to +20 °C, accuracy ±2 °CNone, 4-20 mA, DIN 43650 connectorGet a quote

Gas flow

ModelMethodRangeDisplayQuote this
Laminar differential pressure mass flowOrder specific, per gasBuilt-in backlit displayGet a quote
Thermal mass flow0.014 mln/min to 1,670 ln/min, order specificNone; separate readout module optionalGet a quote

Pressure transmitter — controlling and logging both sides of the stack

ModelMethodRangeDisplayQuote this
PRIGNITZ MikrosystemtechnikSPT-P1A
Piezoresistive, stainless steel thin-film cellVacuum and 600 mbar up to 600 bar, selectableNone, 4-20 mAGet a quote

Pressure switch — a shutdown interlock on its own

ModelMethodRangeDisplayQuote this
HoneywellLE SeriesLER00600TBMNUBAA01
Electromechanical contact (SPST)Set point selectable from 0.24 to 10.3 barNone, switch contactGet a quote

Pressure gauge — reading it on site without power

ModelMethodRangeDisplayQuote this
YAMAMOTO KEIKIPG-A
Bourdon tube (mechanical), JIS B7505-1Order specific; vacuum and compound ranges availableDial and pointer; no power neededGet a quote

Gas leak — a fixed detector watching the area

ModelMethodRangeDisplayQuote this
MSA SafetyULTIMA X5000
Catalytic bead or electrochemicalCatalytic 0–100% LEL or 0–20% LEL; electrochemical 0–1,000 ppmBuilt-in OLEDGet a quote
Catalytic bead or electrochemicalCatalytic 0–100% LEL, full scale settable between 20 and 100%; hydrogen electrochemical 0–1,000 ppmBuilt-in tri-colour backlit LCDGet a quote

Gas leak — a portable detector to find the leak

ModelMethodRangeDisplayQuote this
New Cosmos ElectricXP-3310II
Catalytic bead, automatic pump sampling0–100% LEL, switching to the low range automatically; target gas specified at orderBuilt-in; digits and barGet a quote
Hydrogen-selective sensor, tracer-gas sniffingThe part under test is filled with tracer gas (5% H₂, 95% N₂). Not for watching a room in operationFrom 0.5 ppm H₂ in measuring mode; from 5×10⁻⁷ mbar·L/s in locating modeTouchscreen on the unit, LCD on the probeGet a quote

Reading this table

Hazardous area
Give us the area classification and we shortlist only models certified for it
Calibration and consumables
We point you to the manufacturer or its local service network
Installation
Sample lines and signal wiring agreed separately
Specification basis
Table values follow manufacturer documentation; we re-confirm the current specification before ordering
Order

What to settle before a quotation

Questions we get

Only what has come up more than once in real enquiries.

What to measure

Does hydrogen in oxygen rise when running at part load?

It does, and raising the pressure raises it again. Measured: at ambient pressure, 0.39 percent at rated load and 0.66 percent at 40 percent load; at 6 barg, the same two points are 0.98 and 1.75 percent. Read the multiple rather than the absolute value. Dropping to 60 percent load multiplies it by 1.25 to 1.43 and dropping to 40 percent by 1.67 to 1.94, and those multiples hold across pressures. How much headroom is left is set by pressure. At ambient there is room to spare at 60 percent load against the 2 percent control limit; at 6 barg and 40 percent load only 0.25 percentage points remain. Minimum load under pressure is therefore a function of pressure. For deeper turndown we recommend a minimum load limit, depressurising at low load and a hard interlock on the oxygen side.

Is oxygen in hydrogen the only thing we need to watch?

That value is a purity indicator more than a safety one. The limit is reached first on hydrogen in oxygen. Whether you measure both or only the safety indicator depends on unattended operation and the installation conditions.

Can these readings tell us when to service?

Partly. If hydrogen in oxygen exceeds 2 percent, stop and inspect. That line is set by regulation, so it is not raised for operating convenience. Electrolyte replacement is not decided on this value alone; we look at it together with the cell voltage trend at the same current and the spread between cells.

Which model to choose

We are in a hazardous area. Which model should we take?

If you want to read the markings yourself, see "How to read a hazardous-area marking" above; the reference table of the models we have confirmed is there too. Protection method, gas group, temperature class and equipment protection level are run together into one line, and hydrogen is at the demanding end, so look for IIC or IIB+H₂. Note also that the hazardous-area version is usually a separate model number rather than an option on the same one, so check the part number on the order once more before it goes out.

Can the analyser alarm go to our controller over the network and trip the system?

For monitoring, yes. For a safety trip, no. A network link can be delayed or drop, and a safety trip cannot be late, so wire it directly. Treat a stop command sent over the network as the secondary path. That is why you should check how many volt-free contacts a unit has when you choose it.

Send us the checklist

Fill in only what you know. We ask about the rest.

  • What the measurement is fore.g. safety monitoring
  • Range and accuracye.g. hydrogen in oxygen, a range past 0–2%
  • Signal and integratione.g. 4-20 mA into a supervisory system
  • Country and area classificatione.g. Korea, non-hazardous indoors
Request a quote for this item

Warranty — Manufacturer warranty passes through