← Components & BoP
Downstream

Instrumentation and safety

Gas quality, dew point and flow measurement across the whole plant

Running an electrolyser means measuring gas quality, pressure, flow, level and leaks. We start by setting out what has to be measured where, then confirm range, accuracy, signal type and the certification required in your country before finding a model.

WaterTreatmentCirculationCell / stackDryingAnalysisCompression
Control basis
H₂ in O₂ below 2%, the limit set by Korean and international safety regulation
Measured under
23-cell HXS-2 · 0.1 M KOH · 43–45 °C · ambient to 6 barg · 49 points
Measured
Ambient 0.39 / 0.66% · 6 barg 0.98 / 1.75% (rated / 40% load)
Understand

What this part does and what decides it

Why it matters

The safety limit is reached on the oxygen line, not the hydrogen line. The indicator is hydrogen in oxygen (H₂-in-O₂) and the ceiling is 2 percent. That is not a figure we set: it is the limit Korean and international safety regulation both work to. Sizing the measuring range from the rated point is not enough, because two effects stack. Permeation through the membrane is nearly independent of current while oxygen production scales with it, so at lower load the same permeation is diluted into less oxygen and the ratio rises. On top of that, raising the pressure widens the hydrogen partial pressure difference across the membrane and increases the permeation itself. Measured on a 23-cell HXS-2: at ambient pressure, 0.39 percent at rated load and 0.66 percent at 40 percent load; at 6 barg, the same two points are 0.98 and 1.75 percent. Set range and alarms from the lowest load and the highest pressure you will actually run, together. The method decides the running cost. Principles with no consumable cell need periodic calibration and nothing else; electrochemical cells have a service life, so replacement and a shorter calibration interval come with them. If trace oxygen at ppm level pushes you to an electrochemical unit, put that upkeep in the budget from the start. What each principle measures, and where they divide, is set out under "What each measuring principle actually measures" below.

IN OUR OWN SYSTEMS

Over the 2,400 hour continuous run on a 23-cell HXS-2, hydrogen in oxygen stayed below 2 percent for most of the run, with two brief excursions above it. That run was at 0.3 M KOH, 40 °C and ambient pressure. The sweep across load and pressure is a separate dataset: 49 points from ambient to 6 barg at 0.1 M KOH and 43–45 °C, and the four figures above are its four corners. The electrolyte concentration differs between the two datasets, so read that alongside them when you place them side by side. We do not have a load sweep at the 10 barg rating. Oxygen in hydrogen reached a maximum of 0.10 percent on the 2,400 hour run, and 0.10 percent again over the 294 hour window at 10 barg. If you are adding instrumentation to one of our systems, we work out with you how it ties into the leak detection, E-STOP and interlocks already there.

Background

What each measuring principle actually measures

The Measuring principle column says what the instrument physically responds to. For the same oxygen concentration, each principle is strong over a different range and carries different consumables and calibration intervals. Read this column first and it becomes clear why some models belong at trace level and others at percent level. There are two families. Principles that consume the target gas reach lower, but the sensor has a service life. Principles that do not consume it last longer and hold calibration longer, but lose resolution as the concentration falls. The two flow meters are a different matter altogether: they measure how much is flowing, not what it is made of. How far down a given model reaches varies within a principle. That is what the Range column in the table records.

Electrochemical, fuel-cell type
Measures the current produced as the target gas reacts at an electrode; the current is proportional to concentration. The cell generates its own current like a fuel cell, so no external supply is needed. Strong at low levels, reaching ppm and, on some models, below. The electrode and electrolyte are consumed by the reaction, so the cell is a replacement item
Polymer electrochemical
A solid polymer membrane in place of a liquid electrolyte. Nothing to dry out or leak, so there is no top-up and no restriction on mounting orientation
Thermal conductivity
Uses the fact that gases carry heat at different rates: the heat leaving a warmed element reads the composition. Hydrogen conducts about seven times as well as air, so the contrast is large and this principle suits it particularly well. No consumables, but it only works on a two-component mixture; a third gas leaves the reading ambiguous
Thermo-paramagnetic
Oxygen is drawn into a magnetic field and other gases are barely affected. The strength of that pull gives the oxygen concentration. It does not consume oxygen, so it is long-lived and strong at percent-level concentrations
Optical luminescence
A fluorescent layer emits light when illuminated, and oxygen reaching it suppresses that emission. Concentration is read from how far the emission drops and how long it takes to recover. Oxygen is not consumed, so calibration holds for longer; the fluorescent cap is the periodic replacement item
Thermal mass flow
Measures the heat a flowing gas carries away from a heated element. The correction factor differs by gas, so the unit has to arrive configured for hydrogen
Laminar differential pressure mass flow
Uses the fact that the pressure drop across a laminar element is proportional to flow. Pressure and temperature are measured alongside it and converted to mass flow. Response is fast

How to read a hazardous-area marking

If the instrument goes into a hazardous area it has to be rated for that area. A single line such as Ex db IIB+H₂ T6 Gb in a catalogue is four separate things written end to end. The db at the front is the protection method. Flameproof (the Ex d family) contains an ignition inside a strong enclosure so it cannot spread; intrinsic safety (Ex ia, Ex ib) keeps the energy in the circuit below what it takes to ignite anything in the first place. Analysers with a display and mains power are usually flameproof; small transmitters with no display are usually intrinsically safe. The IIB or IIC in the middle is the gas group. The further along it goes the easier the gas is to ignite, and hydrogen sits in the most demanding group, IIC. A unit certified only to IIB cannot be used in a hydrogen area. Some markings add hydrogen on its own as IIB+H₂, so do not rule a unit out on the letters IIB alone. T6 is the temperature class. It says the surface stays below the ignition temperature of the gas, and T1 through T6 guarantees progressively lower surface temperatures. The Gb at the end is the equipment protection level: Ga corresponds to Zone 0, Gb to Zone 1, Gc to Zone 2. The certification schemes differ by country: ATEX in Europe, IECEx internationally, KCs in Korea. The same unit may still need a separate KCs certificate to be installed in Korea, so tell us the country of installation first. Below is only what we have confirmed against manufacturer documentation. A model left out is not a model without a hazardous-area version; it is one we have not confirmed. Note also that the hazardous-area version is usually a separate model number rather than an option on the same one. Get the part number wrong on the order and a general-purpose unit arrives.

Michell XTC601
Flameproof Ex db IIB+H₂ T6 Gb; ATEX, IECEx, KCs
Michell XTP601
General purpose (GP) and hazardous area (EX) versions
Panametrics XMTCpro
Flameproof Ex db IIC T6 Gb, Ex tb IIIC T78°C Db; ATEX II 2 GD; Class I Zone 1
Messkonzept FTC 320
Non-flameproof; FTC 320 EX is a separate model
Panametrics oxy.IQ
Intrinsically safe option; Class I Div 1, ATEX Zone 0
Southland OMD-507
Non-flameproof; intrinsically safe option available
Analytical Industries GPR-1500
ATEX version selectable
Michell Easidew
Easidew I.S. is the separate intrinsically safe model; IECEx, ATEX, UKCA
STORK DEWCom II
CE marking only in the manufacturer documentation; no hazardous-area rating confirmed
Archigas TCD3000
ATEX Zone 1 Ex db IIC T4/T3 Gb, IECEx. IIC covers a wider gas group than IIB+H₂
Bronkhorst EL-FLOW Select
EX-FLOW is the separate hazardous-area range
Everything else
Models we have not yet confirmed against manufacturer documentation

Where you read the value

The same measurement is read in different places depending on the model. That is what the Display column records. A screen on the instrument is the simplest case. Where the sensor and the screen are one body it is integrated with the sensor; where it is built to drop into a cabinet door it is a panel-mount type. Either way somebody reads the number on site. A model without a screen only puts the value out as a current. That is the 4-20 mA signal, and the number is visible only if something at the other end receives it: a PLC, a recorder, or the manufacturer's own separate readout. Miss this and the instrument is fine while nobody can see what it reads. Not all 4-20 mA wiring is the same. A two-wire loop carries power and signal on the same pair; a three-wire unit takes its power separately. Connectors are standardised too, so a model using a DIN 43650 connector needs the matching socket ordered with it. Instruments with a wide span often add auto-ranging: the unit changes its own measuring range as the concentration moves. On a bare 4-20 mA link that means the same 20 mA is 1000 ppm at one moment and 100% at another. Decide whether you also want the signal that identifies the current range.

Sent is not the same as done

This is where instrumentation most often goes wrong against upper-level control. Whether the command reached the unit, whether the unit accepted it, whether it is running, whether it finished, whether it failed: each of these has to be a separate state. Without that separation you get a screen that reads normal while nothing is actually happening.

Agree units and scaling first

The same quantity comes in different units from different instruments. This table alone has oxygen-in-hydrogen reported in percent and in ppm. If the tag units are not matched and the 4-20 mA scaling is carried over as is, the reading is out by a factor of ten thousand, and it often sits in a plausible range long enough that nobody notices. If you also bring pressure in, settle barg against bara for the same reason.

Choose

The options and the models we handle

What you can choose

Start here and we adjust to your conditions. We can usually source what is not listed.

Safety monitoringBuilding your own stack or system setup

Continuous hydrogen concentration on the oxygen line. That is where the limit is reached first, so unattended operation starts here. The H2 in O2 group below is where these sit.

Quote this option
Hydrogen qualityFuel cells, refuelling, analytical supply

Oxygen in hydrogen and dew point. These are the two values behind a stated purity target. Read the O2 in H2 and Dew point groups below together.

Quote this option
Performance measurementStack evaluation and data

Hydrogen and oxygen flow measured separately and compared with the theoretical rate from current. This is what gives you Faraday efficiency. The Gas flow group below is where these sit.

Quote this option

Pressure, temperature and level are needed to run a stack too, but we do not supply them. Where each one goes is set out in the instrumentation table in the HXS-2 stack integration guide.

Models we can source

Oxygen-in-hydrogen and hydrogen-in-oxygen are measured by different methods, and where you read the value decides the rest. On oxygen in hydrogen the order of magnitude of your target is what selects the model, so settle first whether you are managing it in ppm, as a fuel cell or refuelling specification requires, or whether percent is enough for a purity check. Thermo-paramagnetic is strong across the percent range; electrochemical reaches ppm and, on some models, below. The Range column is that order of magnitude. Hazardous-area rating is not a column here: only the models we could confirm against manufacturer documentation are listed in the table under "How to read a hazardous-area marking" above.

H2 in O2 — watching the safety limit

ModelMethodRangeDisplayQuote this
Michell InstrumentsXTC601EX1
Thermal conductivitySelectable from 0–1% up to 0–100%Built-in touchscreenQuote this
MesskonzeptFTC 320
Thermal conductivitySelectable from 0–0.8% up to 0–100%Built inQuote this
PanametricsXMTCpro
Thermal conductivityOrder specific, per gas pairWith or without, by part numberQuote this
Teledyne Analytical Instruments2000 · 2020
Thermal conductivityOrder specific, per gas pairBuilt inQuote this
Thermal conductivitySelectable from 0–10% up to 0–100%Built-in colour touch panelQuote this
ArchigasTCD3000
Thermal conductivity0 to 5%None, 4-20 mA three-wireQuote this
KNRH2-EC-OEM
Solid-polymer electrochemical0 to 4% H₂None, UART onlyQuote this

O2 in H2 — checking hydrogen purity

ModelMethodRangeDisplayQuote this
Michell InstrumentsXTP601
Thermo-paramagneticSelectable from 0–0.5% up to 0–50%Built inQuote this
Teledyne Analytical Instruments3000MA
Magnetodynamic (dumbbell)Three user-selectable ranges down to 0–1%Built in, auto-rangingQuote this
MagnetopneumaticOrder specific, per module5.7 inch touch screen, built inQuote this
Southland SensingOMD-507
Electrochemical, fuel-cell type0 to 10 ppm through 0 to 100%, selectableBacklit, auto-rangingQuote this
Panametricsoxy.IQ
Electrochemical, fuel-cell type0 to 1000 ppmBuilt inQuote this
Analytical IndustriesGPR-1900
Electrochemical, fuel-cell typeppm level, 0.05 ppm sensitivityPanel-mount typeQuote this
Analytical IndustriesGPR-1500
Electrochemical, fuel-cell typeppm levelIntegrated with sensorQuote this
KNRO2-OEM
Luminescence optical0 to 25% O₂None, UART onlyQuote this

Dew point — checking the hydrogen is dry

ModelRangeDisplayQuote this
Michell InstrumentsEasidewEA2-TX
Dew point −100 to +20 °C, accuracy ±2 °CNone, 4-20 mA two-wireQuote this
STORK InstrumentsDEWCom II
Dew point −100 to +20 °C, accuracy ±2 °CNone, 4-20 mA, DIN 43650 connectorQuote this

Gas flow — measuring output and efficiency

ModelMethodRangeDisplayQuote this
Laminar differential pressure mass flowOrder specific, per gasBuilt-in backlit displayQuote this
Thermal mass flow0.014 mln/min to 1,670 ln/min, order specificNone; separate readout module optionalQuote this

You can buy direct from the manufacturer. Coming through us means we carry the international payment, the customs clearance and the minimum order, and you deal with a domestic invoice.

Order

What to settle before a quotation

What often goes wrong

Only what has come up more than once in real enquiries.

What to measure

Does hydrogen in oxygen rise when running at part load?

It does, and raising the pressure raises it again. Measured: at ambient pressure, 0.39 percent at rated load and 0.66 percent at 40 percent load; at 6 barg, the same two points are 0.98 and 1.75 percent. Read the multiple rather than the absolute value. Dropping to 60 percent load multiplies it by 1.25 to 1.43 and dropping to 40 percent by 1.67 to 1.94, and those multiples hold across pressures. How much headroom is left is set by pressure. At ambient there is room to spare at 60 percent load against the 2 percent control limit; at 6 barg and 40 percent load only 0.25 percentage points remain. Minimum load under pressure is therefore a function of pressure. For deeper turndown we recommend a minimum load limit, depressurising at low load and a hard interlock on the oxygen side.

Is oxygen in hydrogen the only thing we need to watch?

That value is a purity indicator more than a safety one. The limit is reached first on hydrogen in oxygen. Whether you measure both or only the safety indicator depends on unattended operation and the installation conditions.

Can these readings tell us when to service?

Partly. If hydrogen in oxygen exceeds 2 percent, stop and inspect. That line is set by regulation, so it is not raised for operating convenience. Electrolyte replacement is not decided on this value alone; we look at it together with the cell voltage trend at the same current and the spread between cells.

Which model to choose

We are in a hazardous area. Which model should we take?

If you want to read the markings yourself, see "How to read a hazardous-area marking" above; the reference table of the models we have confirmed is there too. Protection method, gas group, temperature class and equipment protection level are run together into one line, and hydrogen is at the demanding end, so look for IIC or IIB+H₂. Note also that the hazardous-area version is usually a separate model number rather than an option on the same one, so check the part number on the order once more before it goes out.

Can the analyser alarm go to our controller over the network and trip the system?

For monitoring, yes. For a safety trip, no. A network link can be delayed or drop, and a safety trip cannot be late, so wire it directly. Treat a stop command sent over the network as the secondary path. That is why you should check how many volt-free contacts a unit has when you choose it.

What we confirm before quoting

Once these are set, selection and quotation happen in one step. If they are not, we start there.

  • What the measurement is for: safety, hydrogen quality or performance
  • Required range and accuracy
  • Signal type and how it ties into your controls
  • Country of installation and the area classification

Supply terms

Supply scope
Model selection, procurement and domestic delivery
Warranty
Manufacturer warranty passes through
Hazardous area
Give us the area classification and we shortlist only models certified for it
Calibration and consumables
We point you to the manufacturer or its local service network
Installation
Sample lines and signal wiring agreed separately
Specification basis
Table values follow manufacturer documentation; we re-confirm the current specification before ordering

Send us the checklist

Fill in only what you know. We ask about the rest.

Request a quote for this item